الوصف
Chandler Pressurized Curing Chambers
Chandler Engineering offers several pressurized curing chambers that are specifically designed to cure standard two-inch cement cube samples for compressive strength testing in accordance with API and ISO standards for oilfield cements. The curing chambers cover a wide range of operational temperatures and pressures to simulate a wide variety of downhole conditions during the curing process.
Selection of the best instrument for your application is based on the number of cubes needed for testing, the size and features of each curing chambers.
Each of the curing chambers has been designed for safe operation and have pressure relief and over-temperature protection. The curing chambers are an important part of any oilfield cement laboratory and are part of a complete line of cement testing instruments offered by Chandler Engineering. The Chandler Engineering Model 4207D Compressive Strength Tester is also an essential instrument used in conjunction with determining the compressive strength of the cube after curing in the curing chamber.
The Chandler Engineering Pressurized Curing Chambers have been used in hundreds of laboratories and have application in research centers, product development laboratories and field laboratories.
Operational Simplicity
The curing chambers are simple to operate. All of the operational controls are conveniently located on the front panel. The temperature and pressure are easily read on the panel gauges and digital indicators.
A programmable temperature controller is capable of controlling multi-slope temperature gradients during a test. In addition, the temperature controller will control the cooling rate at the end of the test in conjunction with the application of cooling water. Pressure is generated with an air-operated high-pressure pump, and control is maintained with a pressure relief valve.
Specifications–
Features- Single or Dual Cell Instruments
- Capacity up to 16 standard cubes
- Temperature control with multi-slope gradient capability
- External chiller connections
Model # Number
of CubesInput
Power
kVAMaximum
Temp PressWeight
Net ShipShipping
Dimensions
W x D x HºF ºC psi MPa lb kg lb kg 1910 16 9.5 700 370 25,000 173 1030 470 1280 580 41x38x79 in
(104x96x200 cm)7355 16 8.5 700 370 5,000 35 780 350 1080 490 41x38x79 in
(104x96x200 cm)7360V 4 or BP
settling
tube test3.2 600 315 6,000 41 150 70 700 320 20x22x19 in
(50x55x48 cm)7370 8 4.5 700 370 3,000 21 520 240 700 320 36x38x66 in
(91x97x167 cm)7375* 16
(8 / cyl)8.5 700 370 3,000 21 1030 470 1200 550 53x38x66 in
(134x97x167 cm)*Model 7375 is a dual cell unit
Operating Temperature: 32 to 120ºF / 0 to 50ºC
Compliance: API Spec. 10A / ISO 10426-1Utilities
Cooling Water: 20-80 psi / 140–550 kPa; nominal flow 2 Lpm
Compressed Air: 100-125 psi / 690-860 kPaPower Supply
220 VAC ±15% 50/60 Hz 7.5 kVA